Micropipette operation and calibration procedure

    1.0 OBJECTIVETo describe a procedure for Calibration of Micropipette by using water.

    2.0 SCOPE

    This procedure is applicable to Calibration of Micropipette and is applicable in Microbiology

    3.0 RESPONSIBILITY

    3.1 Microbiologist /Officer is responsible for the execution. 

    3.2 HOD or designee is responsible for review, effective implementation and compliance to the calibration of micropipette for execution. 

    4.0 PROCEDURE

     



    4.1 Operation

    4.1.1 Set the required volume by rotating color knob clockwise and anti clockwise.

    4.1.2 Fix the micro tip by applying light pressure.

    4.1.3 Press the Plunger and insert the micro tip in the solution to be taken.

    4.1.4 Release the plunger to get the required solution.

    4.1.5 Transfer the solution by pressing the plunger little more than the normal. Press the dispenser knob to remove the used micro tip.

    Note: Use this pipette (100 -1000µl) for volume not less than 100 µl and     not more than 1000 µl Use this pipette (1 -10ml) for volume not less than 1ml and not more than 10ml.

    4.2 Calibration

    4.2.1 Set the volume at 200 µl, 500 µl and 1000µl for micropipette (T -1000µl).

    4.2.2 Follow the operation procedure from point no 6.1.2 to 6.1.5.

    4.2.3 Take the water at room temperature as per set volume and weigh it in a weighing bottle, which is previously dried and weighed.

    4.2.4 Record the reading.

    4.2.5 Repeat the analysis for 5 times and take the average.

    4.2.6Calculate the volume by dividing the average weight of water with the actual volume from the table given as per the test temperature and record the values in Attachment #1 

    Temperature

    Volume of 1 gm of water

    Temperature

    Volume of 1 gm of water

    15 0 c

    1.0020 ml

    24 0 c

    1.0037 ml

    16 0 c

    1.0021 ml

    25 0 c

    1.0039 ml

    17 0 c

    1.0023 ml

    26 0 c

    1.0042 ml

    18 0 c

    1.0025 ml

    27 0 c

    1.0045 ml

    19 0 c

    1.0026 ml

    28 0 c

    1.0047 ml

    20 0 c

    1.0028 ml

    29 0 c

    1.0050 ml

    21 0 c

    1.0030 ml

    30 0 c

    1.0053 ml

    22 0 c

    1.0032 ml

    31 0 c

    1.0056 ml

    23 0 c

    1.00345 ml

    32 0 c

    1.0059 ml


    4.2.7  Acceptance criteria:

    For micropipette 100µl –1000µl

    Set volume in          Permitted Volume         

    200-µl         196-204 µl

    500-µl          495-505 µl

    1000-µl          990-1010µl

    Frequency – Quarterly

    4.2.8 Precautions :

    4.2.8.1 After fixing volume of micropipette don’t rotate the push button.

    4.2.8.2  No foreign particle should be present in the tip as this can block the delivery of the liquid and this can result in incorrect volume dispenses.